Adsorbing cartridges
One of the performance characteristics of radiello® sampling systems specified by EN ISO 23320 (for workplace air) and UNI EN 13528 (for ambient air) is the blank value, i.e. the mass of the target substances detected in new cartridges. The blank value is determined by analysis on random sampled cartridges of each production batch.
Testing is carried out immediately after completion of the production process. The batch is released for shipment only if the masses of all target compounds to be quantified in the cartridges are below the predetermined acceptance criteria. These values have been set to comply with the requirements of EN ISO 23320 and UNI EN 13528, as a minimum.
The raw materials used in the production of radiello® diffusive samplers are subject to incoming quality control to verify their compliance with pre-established requirements.
The finished products (except those that require simple assembly and packaging operations) undergo different quality control checks for each production batch. These checks may include:
The background (“blank”) contamination level of the adsorbing cartridges.
The amount of substance in the preloaded cartridges or calibration solutions, supplied as sampler accessories.
The consistency of the morphological characteristics and the integrity of the sampler components and accessories.
Raw materials are released into the production process and finished products are transferred to the shipping warehouse only after the relevant quality control checks have been successfully completed.
Diffusive body QA/QC
Diffusive bodies are essential components of radiello® sampling systems. Sampling rates depend on the physical and geometrical characteristics of the diffusion membranes: porosity, internal and external diameter, and length.
To ensure the accuracy and reproducibility of the sampling rate, it is necessary to verify that the physical and geometrical characteristics of the diffusion membranes remain unchanged over time and reproducible from batch to batch.
The quality control process involves two stages:
- verification that the manufacturer’s quality controls on each batch of membranes meet the product requirements;
- additional testing in our laboratories of the membranes’ physical and geometrical characteristics, followed by verification of compliance with the requirements for use.
From each batch, a pre-set number of pieces are randomly extracted for quality control purposes. The batch is released for the production of diffusive bodies only if all the physical and geometrical characteristics meet the specified requirements.
Compliance with these requirements ensures that the performance specified in the technical data sheets for radiello® adsorbing cartridges is consistently achieved when the diffusive bodies are used in the correct combination with the selected adsorbing cartridge.